Overview

We have mesoSPIM v5 lightsheet designed by Fabian Voigt in the Helmchen Lab, Zurich. The system is based around an Olympus zoom macroscope and is ideal for screening large numbers of samples and visualising sparse cell populations, plaques, or blood vessels. The system is unsuited for samples where resolutions of about a micron or better are needed. Images acquired on a Hamamatsu ORCA Flash 4.0. Two iChrome MLE laser launches provide excitation wavelengths of 488 nm, 561 nm, 640 nm, and 685 nm. Emitted light is filtered using a 10 position Ludl filter wheel. Open source Python code runs the microscope.

Basic Usage

Turning on the system

  1. Turn on Hamamatsu camera with switch at the back of it.
  2. Turn keys of each laser launch clockwise by 90 degrees. Each laser will play a little tune once it's started up.
  3. Start the mesoSPIM software via the icon on the Desktop.

Turning off the system

  1. Turn off camera.
  2. Switch off lasers with key switch.
  3. Close the software.
  4. Cap the large cuvettes so dust does not get in.

System Details

Filters

Pos. Use Filter ID
1 Quad line ZET405/488/561/640mv2
2 GFP FF01-520/35
3 tdTomato & mCherry BLP01-594R
4   BLP01-633R
5 700 nm low pass  
...    
10 open port n/a

The following filters are available but not installed in the wheel: LP02-561RU, ET585-40m, FF01-565/24 (Could be good for AF555), FF01-515/LP, FF01-542/27, ET535-30m, FF01-530/43, FF01-509/22, FF01-482/35.

Zooms

The MesoSPIM uses an Olympus macroscope zoom body. A stepper motor turns the zoom knob via the software. The zooms correspond to the following pixel sizes.

Zoom FoV Pixel Size
0.63 21.56 0.0105
0.80 16.86 0.0082
1.00 13.42 0.0066
1.25 10.79 0.0053
1.60 8.36 0.0041
2.00 6.69 0.0033
2.50 5.34 0.0026
3.20 4.16 0.0020
4.00 3.29 0.0016
5.00 2.62 0.0013
6.3 2.12 0.0010

The NA varies with Zoom: NA = 0.075 at 1× zoom and NA = 0.22 at 4× zoom. The change in NA keeps pace with magnification so the image does not dim up to about zoom 4. Beyond zoom 4 the NA no longer increases so the image dims. NA 0.22 corresponds to a lateral resolution of about 1.4 microns.