Skip to content

Leica SP8 Confocal

Rob Campbell edited this page May 1, 2024 · 6 revisions

Features

  • Location: Room B273
  • Serial No: 8100001591
  • Motorised XY Scanning Stage Operation with software extension for tile scanning
  • Epi fluorescence (CCD Camera) Operation
  • Laser Set: Super Continuum “White Light Laser” (470-670nm) + 405 & 442nm CW
  • Linear Unmixing (Available during acquisition – Real time fluorescence unmixing
  • Configurable USB Control Panel
  • Three detectors: 2x HyD and 1x PMT. The HyDs can be used for photon counting.
  • Deconvolution option via Huygens.

Overview

The SP8 has a white light laser and tunable bandpass filters. This means it can tune to create whatever excitation wavelength is needed and also catch whatever emission of light you need. There are no set laser wavelengths or filter sets to work around, like on the Zeiss Imager and Axioscan.

Objectives

Magnification Numerical Aperture Field of View Working distance Resolution at 1024x1024(/pixel) Link for more info
x10 0.4 1.11mm x 1.11mm 2.74 mm 1.08um 506407
x10 water immersion 0.4 1.11mm x 1.11mm 360um   506293
x25 water immersion 0.95 442um x 442um 2.4 mm 432nm 506375
x40 oil immersion 1.3 276um x 276um 240 um 270nm 506358
x63 oil immersion 1.4 175um x 175um 140 um 171nm 506350
x20 Clearing Obj (BABB) 0.95   1.95 mm   507702
x25 Motor Correction Collar (CLARITY) 1   6 mm   507703

Important

  1. Move the sample stage/arm out of the way from objectives both before/after your session.
  2. Use the correct immersion medium for the objective
  3. Turn off the HXP lamp if there will be no user within 2 hours.
  4. Contact Peter for training or troubleshooting

Basic Usage Guide

Software

The LAS AF software is comprised of four domains.

  1. Image Acquisition & Control Settings: these functions control the static acquisition settings for image collection, such as Frame Size, Scan Speed, Averaging, Zoom, etc.
  2. Beam Path Configuration: where to set the beam path configuration and control the detector settings to optimize the fluorescence signal.
  3. Image Display: this area is where the image will be displayed. Settings here allow the user to control how the image is viewed on screen.
  4. Scan Action Functions: these functions start/stop scanning and initiate our experiment acquisition.

Mounting Sample on Microscope

Once the system has been powered on, we can mount our specimen on the stage below the microscope objective.

  1. Raise stage to appropriate height and mount sample.*need picture about home and focus bottons
  2. Switch microscope TFT to ‘FLUO’ to activate the incident path from default ‘CS’ in Combi.
  3. Choose filter.
  4. Open/Close reflected light shutter with ‘IL-Shutter’.
  5. Via the microscope eyepieces, visualize your sample and adjust the focus (Z) and stage position (X,Y) accordingly
  6. When sample is in focus and positioned correctly – switch microscope back to CS – Combi to begin imaging with the confocal.

Beam Path Configuration

  1. Open the dye assistant
  2. Choose your fluorochromes from the dropdown list.
  3. Define which type of detector you need.
  4. For the least bleedthrough, choose the line sequential setting.
  5. Be patient for the hardware to initialize!

Check Your Sequential Scan

  1. The SEQ will indicate that you are scanning sequentially.
  2. Click between the different sequences (without scanning) and check that emission bands are not moving between sequences.

Acquisition Settings

The Acquisition Mode parameters located on the left-hand side of the software display contain all the settings available for image optimization and collection.

  1. From the dropdown, choose the Acquisition Mode. For example, if you want to collect 3d image stacks, select ‘xyz’ from the dropdown
  2. The remainder of the settings in the acquisition mode control dialog can adjusted, depending on sample morphology and signal quality.

Image Format

The default image size in pixels is 512x512, however this can be optimized based on zoom and objective settings. For "ideal" Nyquist sampling you can quickly select with ‘Optimize XY Format’. However, this is not always necessary and is use-case dependent. Ask us if unsure.

The default pixel depth is 8 bits If you want to change that, see the "Configuration" tab, "Hardware" submenu, resolution drop down list.

Speed

  1. Scan speed is measured in Hz (lines per second). It refers to the frequency of two galvanometers which drive scanning mirrors, shading laser light across the specimen.
  2. The slower the speed, the higher singal-to-noise ratio (SNR). You can select between 10-1800 H, but 600H is recommended value.
  3. Bidirectional scanning will double the speed as pixels are recorded in both scan directions.
  4. Sometimes when bidirectional is selected, mismatch in the phase might occur, you can correct this with the control panel (phase correction)

Averaging

To improve your image quality, you might choose to apply averaging. You can choose between line averaging, frame averaging or a mixture of both. Averaging takes the sum of pixels from the specified number of scan and uses arithmetic mean as the final value of the image. Averaging preserved those persistent pixel values that are mostly specific signal while divide away along times those fluctuated values that are mostly noise.

Troubleshooting

If you see the message "Safety interlock activated - no laser light. Please check position of safety interlock" you should turn the key on the box under the microscope.

Clone this wiki locally